OMNIACERA1000series
|
|
|
|
|
RA1100
|
RA1200
with built-in printer
|
RA1300
with high density & high speed built-in printer
|

Affordable price.
We offer the main unit with basic functions at affordable price
and make various optional functions to be added on demand. |
High-speed chart feed
The real-time printing function, which is one of the basic
functions of the recorder,features a performance of up to 100
mm/s (RA1300). For data backup, simultaneous data filing is also
available. |
Necessary functions can be selected.
Extended functions (arithmetic operations, FFT analysis, and
waveform judgement) and interfaces (RS-232C, GP-IB, and SCSI)
are provided as optional functions. You can select the functions
according to the measurement conditions. |
Color LCD display and easier operation
The 10.4-inch TFT color LCD display and touch panel support
easier operation.The 10.4-inch TFT color LCD display and touch
panel support easier operation.Setup conditions can be changed
by simply touching the screen. |
Multi-channel support.
16-channel measurement is enabled in this compact unit by the
newly developed dual-amp (2 chs./unit). |
Long-time data filing
Long-time data acquisition (filing) is possible at a sampling
speed of 200 s by storing data in the optional built-in MO (640
Mbytes) or PC card. Since the recorded data is in a digital
format, data analysis after recording and long-term data
management, which were impossible with chart data, can be
performed. |
Variety of input units.
16-bit high-resolution amps, which support the input of voltage,
vibration, strain, and rotation pulses, are provided. In
addition a 1µs (1MS/s) sampling speed is available by using a
12-bit high-speed DC amp. Replacement of input units is easier
since they feature a plug-in design. |
FAX
transmission function increases measurement efficiency.
Automatic data transmission to a fax machine during data
acquisition is possible by connecting the recorder to a modem. |
 |
|
|
TC/DC
amp (AP11-107)
This amp can be used as a thermocouple amp or DC amp. When
temperature measurement using a thermocouple amp and voltage
measurement using a DC amp are necessary, this amp is useful
because no amp replacement is required.Moreover, four types of
thermocouples, R, T, J, and K, can be connected and high-speed
temperature measurement as fast as 10 µs is possible. This amp
contains one channel per unit.
Event amp (AP11-105)
This amp is a logic amplifier that can input eight channels of
voltage or contact signals. It is useful for measuring timing or
contact signals.
Voltage input: Input voltage 0 to 24 V
H level: 2.5 V or higher
L level: 0.5 V or lower
Contact input:
Open: 2 k
or higher
Short: 250
or lower
2-ch AC strain amp (AP11-104)
This amp is a 2-channel amplifier that inputs signals from
strain gauges or straingauge based sensors. Due to the AC bridge
type, it features high noise resistance(against the noise caused
by the line voltage), high stability, and a high resolution of
16 bits. Low-noise measurement is possible for measurement using
strain gauges, which do not have any shielded wires.
Auto-balance
Bridge voltage: 5 kHz, 2 V rms
Response frequency: 0 to 2 kHz
Gauge factor adjustment function: 1.9 to 2.2
2-ch DC strain amp (AP11-110)
This amp is a 2-channel amplifier that inputs signals from
strain gauges or strain gauge based sensors. Due to the DC
bridge type, it features measurement of high-speed strain
signals at a high resolution of 16 bits.
Auto-balance
Bridge voltage: 2 VDC, 5 VDC
Response frequency: 0 to 50 kHz
2-ch vibration/RMS amp (AP11-109)
This is a 2-channel amplifier that features a constant-current
power supply for amp-embedded sensors, an RMS conversion
function, and a band-pass filter.The RMS function in this amp is
effective for monitoring the basic frequency level fluctuation
in rotation equipment and the fluctuation of the AC power
level.For vibration measurement, an amp-embedded charge sensor
can be directly connected to this amp unit. For voltage
measurement, 500 V (DC or AC peak values)can be directly input.
In addition, the band-pass filter offers filtering of necessary
frequencies to perform the RMS conversion so that clear
waveforms for judgment are available.
Power for sensor: 2 mA/18 V or higher
Voltage measurement range: 100 mV to 500 V
Band-pass filter: -24 dB/oct
A/D: 16 bits
2-ch FFT amp (AP11-102)
This is a 2-channel amplifier that incorporates an anti-aliasing
filter required for the FFT processing and the power for an
amp-embedded sensor (constant current).The anti-aliasing filter
has a dropout feature at -72 dB/1.5 fc so that the aliasing
generated in FFT can be eliminated.
Power for sensor: 2 mA/18 V or higher
Voltage measurement range: 100 mV to 500 V
A/D: 16 bits
Fmax: 40 kHz
F/V converter (AP11-108)
This unit offers frequency-to-voltage conversion, which is
useful in recording rotation changes and revolutions in rotation
equipment.
Input frequency: 10 kHz max.
Response time: 20 ms (Range of 10 kHz) |
2-ch
high-resolution amp (AP11-101)
This is a DC amp that can record at a high resolution of 16
bits. Channels are isolated from each other and have a withstand
voltage for input of 500 V max. (DC or AC peak values). This amp
contains two channels per unit.
|
2-ch high-speed DC
amp (AP11-103)
This is a DC amp that can acquire data at a high-speed sampling
rate of 1 µs.Channels are isolated from each other and have a
withstand voltage for input of 500 V max. (DC or AC peak
values). This amp contains two channels per unit.
|
2-ch TC/DC amp (AP11-106)
This amp can be used as a thermocouple amp or DC amp. When
temperature measurement using a thermocouple amp and voltage
measurement using a DC amp are necessary, this amp is useful
because no amp replacement is required.Moreover, five types of
thermocouples, R, T, J, K, and W can be connected and high-speed
temperature measurement as fast as 10 µs is possible. Channels
are isolated from each other and have an input voltage of 50 V
max. (DC or AC peak values). This amp contains two channels per
unit.
|
 |
|
Flaw Detection in
Conveyor Line in Steel Mill |
Operation Test of
Power Switchboards and Production Facilities |
By using the filing function, which operates
synchronously with external signals, data can be acquired for a
long term. Since the RA1000 Series is small and compact, it does
not occupy a large measurement space. |
The RA1000 Series is used to measure the
operation timing of relays and voltage and current changes in
the main power for operation. |
|
Torque Measurement of Engines
or Motors |
Wide Variety of Tests in
Railroad Car |
In automobile testing, measurement of
revolution, strain, pressure, vibration, torque, temperature,
etc is executed. The RA1000 Series incorporates a variety of
amplifiers, which can eliminate the need to install separate
amplifiers and thus enable operators to perform the tests in a
small space.
By using the filing function of the RA1000 Series, long-term
data storage in external media at a high sampling speed (200 µs
max.) is possible. |
The RA1000 Series can accept oil pressure, air
pressure, and acceleration in its AC strain amp, and save the
data in the MO directly. Since the RA1000 Series incorporates
strain amplifiers, external amplifiers are not necessary, which
makes the measurement system compact.
Also, a long-term multi-channel measurement is possible by using
the filing function.
|
|
Governor Test |
Drop Test for Package |
The RA1000 Series is used as a recorder for the
governor test, in which changing signals such as water pressure,
revolution, generator voltage, current, governor, stroke, and
pressure regulation stroke are recorded.
Since the data saved in the memory can be automatically stored
in a PC card, database creation is possible without expending
excessive time and effort. After storing the data, you can
confirm the data on the monitor, which eliminates errors caused
by variations in reading by operators. |
A drop test for packages is performed in order
to avoid the damage that may occur during delivery of packaged
electronic equipment.
Generally, packages contain materials for packing to absorb
shock. This test measures the impact acceleration to obtain the
data that is used for the improvement of materials for packing.
There are two types of accelerometers that can be used in the
drop test: a piezo-electric type or a strain gauge type. The
RA1000 Series supports both types. |
|
Monitoring of Vibration Level
of Rotating Machine |
Magnus Test |
In order to maintain rotating
equipment such as motors, gears, bearings, piezo-electric
accelerometers are used to check their vibration level.In this
case, a band-pass filter that picks up the signals in the
vicinity of the frequency of the rotation is used and RMS
conversion is performed based on the signals so that the level
changes become more visible.Waveform judgment setting is also
possible. |
A Magnus test is a test to examine
the response of a sample of tissue that expands or contracts,
such as the blood vessel, enteron, or trachea of a mouse or rat,
to the medicine applied to the sample.
This test requires a lot of repetition work. The filing and
trigger functions of the RA1000 Series provide automatic
processing of this work.Moreover, since strain amps can be
incorporated, a tension sensor can be directly connected to the
main unit,thus eliminating troublesome cabling.In addition, an
operator can examine waveforms on the monitor during data
acquisition and readout current values calculated by physical
conversion. |

Basic Specifications
| Display Block |
10.4-inch TFT color LCD
Effective screen area: 211.2 mm x 158.4 mm (640 dots x 480 dots)
|
| Printer Block (RA1200/RA1300) |
| |
Printing method |
Thermal printing using a thermal head |
| |
Paper Width |
219.5 mm |
| |
Effective Recording Width |
1 split (200mmFS) to 16 splits (10mmFS), Number
of split and
printing width can be changed. |
| |
Channel Discrimination |
Prints channel number in the vicinity of the
printed waveform
The ON/OFF function is available. |
|
Battery Backup |
Backup Contents: Main unit settings,
Recording conditions
Backup time: About 1 month (if fully charged and used under
normal temperature) |
|
Storage Devices |
A built-in 3.5-inch floppy disk drive
(1.25MB/1.44MB)
PC card slot (ATA flash memory card, SRAM card)
Built-in MO unit (640, 540, 230, and 128MB supported): optional
|
|
Interface |
RS-232C unit (38400 to 2400 bps): optional
GP-IB unit (IEEE488 compliant): optional
SCSI unit (50-pin type): optional |
|
Remote Terminal |
Start/stop, mark print, and paper feed are
available by external signal,
synchronous pulse Input, protect input, output of waveform
judgement
results, and error output |
|
Clock Function |
Stability: Daily error within ±35 PPM (at normal
temperature) |
|
Operating
Environment |
Temperature: 0 to 40°C (5 to 40°C
when FFD is operating)
Humidity: 35 to 80% RH (without condensation)
Vibration Resistance: MIL-STD-810E compliant (except when FD or
MOis operating) |
|
Power Supply |
90 to 132 VAC/180 to 264 VAC (to be specified
when ordering)
Frequency 47 to 63Hz
DC power unit (optional) In use: 11 to 28 VDC(AC power is
prioritized) |
|
Power
Consumption(Maximum) |
RA1100: Approx. 85 VA
RA1200: Approx. 180 VA (About 85 VA during stand-by)
RA1300: Approx. 350 VA (About 85 VA during stand-by)
|
|
Dimensions |
Approx. 369.5 (W) x 150.5 (H) x
301(D) mm |
|
Weight |
Approx. 6.6kg (RA1100 main unit with
four 2-ch high-resolution DC amps)
Approx. 7.5kg (RA1200 main unit with four 2-ch high-resolution
DC amps)
Approx. 7.5 kg (RA1300 main unit with four 2-ch high-resolution
DC amps) |
Trigger Specifications
|
Trigger Mode |
OR, AND, A*B, WINDOW |
|
Trigger Source |
Input Signal, time trigger (starting
time, ending time, measurement
interval), manual trigger and external trigger |
|
Trigger Setting |
Amps other than event amps
| |
Trigger Slope |
OR, AND..... or |
| |
|
A*B... or for
each A,B source |
| |
|
WINDOW...OUT or IN |
Level Setting: To be set with physical values (e.g. voltage) |
Event Amp
| |
State Settings:
|
H, L, or OFF can be set for each input
(without trigger
slope) When OFF is set, trigger condition is notapplied.
|
| |
Trigger
Setting: |
AND or OR of the state
setting conditions of inputs
from 1 to 8 |
|
|
Trigger Related Functions |
Pre-trigger , trigger pass count,
trigger filter, and trigger mark |
Other Specifications
|
Print Functions
(RA1200/RA1300) |
| |
Data Information |
Measurement mode, year/month/day,
Measurement start time, Data No.,
Trigger conditions (Trigger point, Trigger year/month/day,
Trigger time),
Sampling speed, Chart speed, Time axis, etc. are printed
simultaneously
with waveform print). ON/OFF function is available. |
|
Channel Information |
Input unit settings are printed
simultaneously. ON/OFF function is available. |
|
Mark print |
In the real-time mode, the mark
print is available (M date/time) |
|
List Print. |
Date, time, measurement mode, data
No. , record form, chart speed,
sampling speed, scale printing, back-light auto-off, auto start,
buzzer
click, trigger mode, source channel, recording operation, filing
setting printout |
|
Screen Copy Function |
Prints hard copies of the contents
shown on the screen |
|
A4 report output |
Registers free comments and prints
them together with the waveform printing |
|
Printing Line Width |
Sets baseline boldness (1, 2, 3, 4
dots) for each channel |
|
Key Lock Function |
Invalidates key input to prevent
erroneous operation |
|
Storing and
Reading of Settings |
Up to four conditions for input unit
and main unit settings are stored in
the main unit memory |
|
Physical Value
Conversion |
Physical conversion of input signal,
full-scale modification of display |
|
Stand-by Function |
When the main unit is restored from
blackout or temporary halt, original
state is regained.(Except when filing) |
|
Automatic Setup
Function |
By presetting a floppy disk
containing the measurement conditions in FD
drive, the conditions are automatically read when turning on the
power. |
|
Storing the Screen
Image |
Stores a screen image in bitmap
format into a floppy disk |
|
Help Function |
Displays the handling method,
cautions, etc. |
|
RS-232C Unit
Function |
| |
Remote Control |
Controls the main unit from a personal computer
via the RS-232C cable. |
| FAX Transmission Function |
Sends measured waveform by using a FAX modem
connected to RS-232C. |
|
GB-IB Unit Function |
| |
Remote Control |
Controls the main unit from a personal computer
via GB-IB cable. |
|
SCSI Unit Function |
| |
Stores data to MO or PD as external memory
devices. |
AC Bridge Power Supply Unit (Optional)
|
Function |
Bridge power source for 2-ch AC
strain amp |
|
Power Voltage, Carrier Wave |
2 Vms, Sine wave 5kHz |
|
Synchronism |
Synchronization with other RA1000s
using a built-in AC bridge power unit is available. |
Memory Mode
|
Measurement
Operation |
Once, Repeat, Endless |
|
Memory Capacity |
256kW/CH (Max.1 MW/CH optionally
possible)
*When function operation is used, memory capacity is 1/4
When two channels are used, 2MW/CH is the maximum |
|
Memory Division |
1, 2, 4, 8, 16, 32, 64, and 128
divisions |
|
Sampling |
1 µs (1 MS/s) to 1 s(1 S/s), User-settableExternal
synchronization enabled |
|
Display (Replay
Monitor) |
| |
Waveform Display |
Display magnification :x5 to x1/1000 |
|
X-Y Display |
X-Y display where data of an any
channel is shown on the X axis, and
other channel on the Y axis. Display step: 1 to 1000 steps
|
|
Print
(RA1200/RA1300) |
| |
Waveform Print |
Print Density: Voltage axis: 8
dots/mm Time axis: 10 dots/mm
Copy magnification: x5 to x1/1000 |
| Alphanumeric Print |
Readout Interval : 1 to 1000 steps |
| X-Y Print |
X–Y print where data of an any
channel is shown on the X axis, and other
channel on the Y axis. Effective print range : 200 mm x 200 mm
Print
density : 400 dots (X axis) x 400 dots (Y axis) Display step : 1
to 1000 steps |
|
Memory Filing |
Each time data is stored into memory,
the data is saved in the memory
device in the binary or CSV format. |
Real-Time Mode (RA1200/RA1300)
|
Waveform Printing |
| |
Measurement Starting Operation |
Starts with the Start key, by the
trigger detection, or at set time.
Interval recording available. |
|
Paper Speed |
25 mm/s to 1 mm/sec (RA1200), 100
mm/s to 1 mm/min (RA1300),
User settable. External synchronization enabled |
|
Frequency Response |
DC to 100 kHz (sampling points: 10
points/cycle), Varies depending on the input units |
|
Printing Density |
Voltage axis : 8 dots/mm
|
Time axis |
:RA1200 |
:10 dots/mm |
| |
:RA1300 |
:20 dots/mm at 100 mm/s |
| |
|
:40 dots/mm at 50 mm/s |
| |
|
:80 dots/mm at 25 mm/s |
|
|
Printing Length |
Continuous or 1 to 1000 division-shot
feeding |
|
Alphanumeric
Printing |
| |
Sampling |
1 s to 1 hour, User settable,
External synchronization enabled |
| Printing Length Setting |
Continuous or 1 to 1000 data shot feeding |
|
X-Y Printing |
| |
Performance |
X-Y recording where data of an any
channel is shown on the X axis, and
other channel on the Y axis. Continuous recording until stop
button is
pressed (pose function available) |
| Effective Printing Range |
200 mm x 200 mm |
| Printing Density |
400 dots (X axis) x 400 dots (Y axis) |
| Printing Speed |
10 ms or more |
|
Backup Filing |
Stores the measured data in a memory
device in real-time
Recording Speed: Synchronized with paper speed
|
Transient Mode
|
Transient
Recording (RA1200/RA1300) |
| |
Function |
Normally operates in the real-time
mode. When trigger is detected, the
mode is switched to the memory mode. |
|
Real-time Section |
In conformance with the waveform
printing in the real-time mode |
|
Memory Section |
In conformance with data storage and
printing in the memory mode |
|
Transient Filing |
| |
Function |
Normally operates in backup filing,
data is saved in a memory when trigger is detected, and when
measurement finishes, the memory data is saved. |
| Real-Time Section |
In conformance with backup filing in the
real-time mode |
| Memory Section |
In conformance with memory filing in
the memory mode |
Filing Mode
|
Data Recording |
| |
Function |
Stores the measured data real-time to
a memory device (Real time filing) |
|
Recording Speed |
200 µs to 60 s. Maximum recording
speed varies depending on number
of channels and recording device. User settable External
synchronization ennobled |
|
Recording Method |
Select normal or ring
Ring recording is possible (Repeated recording during set time)
|
|
Waveform Printing
(RA1200/RA1300) |
| |
Function |
Prints the waveform of input signal |
| Paper Speed |
25 mm/s to 1 mm/m (RA1200), 100 mm/s to 1mm/m
(RA1300) User
Settable External synchronization enabled |
| Frequency Response |
DC to 100 kHz (sampling point: 10
points/cycle), Varies depending on
the input units |
| |
Printing Density |
Voltage axis : 8 dots/mm
|
Time axis |
:RA1200 |
:10 dots/mm |
| |
:RA1300 |
:20 dots/mm at 100 mm/s |
| |
|
:40 dots/mm at 50 mm/s |
| |
|
:80 dots/mm at 25 mm/s |
|
FFT Unit (Optional)
|
1 Signal Analysis |
Linear spectrum, power spectrum, RMS spectrum,
power spectrum
density, and octave analysis (1/1, 1/3) |
|
2 Signals Analysis |
Transfer function, coherence
function, and cross-power spectrum |
|
Window Function |
Rectangular, hanning, hamming,
exponential, force |
|
Addition Processing |
Simple additional average (time
axis), simple additional average
(frequency axis), exponential weighted average (frequency axis),
and peak hold |
|
Analysis Frequency |
400kHz, 200kHz, 80kHz, 40kHz, 20kHz,
10kHz, 5kHz, 2kHz, 1kHz,
500Hz, 200Hz, 100Hz, 50Hz, 20Hz, and 10Hz |
|
Resolution |
400 (1000), 800 (2000), 1600 (4000)
Numbers in parentheses indicates the number of sampling points.
|
Arithmetic Operation Unit (Optional)
|
Sectional
Statistical Operation |
Maximum value, minimum value, P-P value, average
value, area, effective value, standard deviation, rising time,
falling time |
|
Function Operation |
Four-function calculation, absolute
value, differential, integration, secondary differential, double
integration, square root, exponential, common logarithm, moving
average, and trigonometric functions (sin, cos,tan, asin, acos,
atan) |
Waveform Judgment Unit (Optional)
|
Function |
Measures waveform of input signals of eight
channels, and outputs the results of waveform judgement |
|
Targets of
Outputting Results |
Print on paper, FAX, data storage,
or remote terminal output |

2-CH High Resolution Amp (AP11-101)
|
Input |
2channels/unit, isolated unbalanced input,
isolated BNC connector |
|
Input coupling |
AC and DC coupling |
|
Input impedance |
1M or
higher |
|
Measurement range |
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,
100, 200, and 500 V |
|
Range accuracy |
Within 0.3% FS (Within 0.8% FS at
500 V) |
|
Offset accuracy |
Within 0.3% FS (at 25°C) |
|
Linearity |
Within 0.1% FS |
|
Allowable input
voltage |
Range of 10 V to 500 V: 500 V max.
(AC or DC peak values)
Range of 0.1 V to 5 V: 100 V max. (AC or DC peak values) |
|
CMV |
Unit only: 43 V (AC or DC peak
values)When using isolated BNC cable
(optional): 300 VAC |
|
Frequency response |
At DC coupling: DC to 50 kHz (+0.5,
-3 dB)
At AC coupling: 0.3 to 50 kHz (+0.5, -3 dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -12
dB/OCT)
30, 300, 3 kHz, and OFF (+0.5, -3 dB) |
|
A/D converter |
16 bits, 100 kHz max. (two channel
simultaneous sampling) |
|
Temperature
stability |
Zero point: Within 0.03% FS/°C
Gain (range): Within 0.01% FS/°C |
|
Weight |
230 g |
2-CH High Speed Amp (AP11-103)
|
Input |
2channels/unit, isolated unbalanced input,
isolated BNC connector
|
|
Input coupling |
AC and DC coupling |
|
Input impedance |
1M or
higher |
|
Measurement range |
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,
100, 200, and 500 V |
|
Range accuracy |
Within 0.5% FS (Within 0.8% FS at
500 V) |
|
Offset accuracy |
Within 0.5% FS (at 25°C) |
|
Linearity |
Within 0.2% FS |
|
Allowable input
voltage |
Range of 10 V to 500 V: 500 V max.
(AC or DC peak values)
Range of 0.1 V to 5 V: 100 V max. (AC or DC peak values) |
|
CMV |
Unit only: 43 V (AC or DC peak
values)When using isolated BNC cable
(optional): 300 VAC |
|
Frequency response |
At DC coupling: DC to 400 kHz (+0.5,
-3 dB)
At AC coupling: 0.3 to 400 kHz (+0.5, -3 dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -12
dB/OCT)
50, 500, 5 kHz, and OFF (+0.5, -3 dB) |
|
A/D converter |
12 bits, 1 MHz max. (two channel
simultaneous sampling) |
|
Temperature
stability |
Zero point: Within 0.03% FS/°C
Gain (range): Within 0.01% FS/°C |
|
Weight |
240 g |
2-CH TC/DC Amp (AP11-106)
|
Input |
2channels/unit, isolated unbalanced input,
terminal block 4
|
|
Input coupling |
DC coupling |
|
Input impedance |
10M or
higher |
|
Thermocouple |
R, T, J, K, and W |
|
Measurement range |
R: 0 to 1600°C
T: -200 to 400°C
J: -200 to 1000°C
K: -200 to 1350°C
W: 0 to 2300°C |
|
Voltage
measurement range |
100, 200, 500 mV, 1, 2, 5, 10, 20
and 50 V FS |
|
Range accuracy |
Temperature: Within 0.5% FS (Within
1% FS at 0°C or lower)
Voltage: Within 0.5% FS |
|
Cold junction
compensation |
Internal/external switchable
Accuracy: Within 2°C (Within 1°C at stable temperature of 20°C
at input terminal) |
|
Linearity |
Within 0.1% FS |
|
Allowable input
voltage |
50 V (AC or DC peak values) |
|
CMV |
43 V (AC or DC peak values) |
|
Frequency response |
DC to 40 kHz (+0.5, -3 dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -18
dB/OCT) 1, 30, 500, 5 kHz, and OFF (+0.5, -3 dB) |
|
A/D converter |
15 bits, 100 kHz max. (two channel
simultaneous sampling) |
|
Temperature
stability |
Accuracy: 0.04% FS/°C (When used as
temperature amp)
Zero point: Within 0.03% FS/°C (When used as DC amp)
Gain (range): Within 0.01% FS/°C |
|
Weight |
240 g |
TC/DC Amp (AP11-107)
|
Input |
1 channel/unit, isolated unbalanced input, 2
binding posts |
|
Input coupling |
DC coupling |
|
Input impedance |
10M
or higher (1M
at 5, 10, 20, 50 V FS in DC amp) |
|
Thermocouple |
R, T, J, K |
|
Measurement range |
R: 0 to 800°C at 800°C FS and 0 to
1600°C at 1600°C FS
T: -200 to 200°C at 200°C FS and -200 to 400°C at 400°CFS
J: -200 to 200°C at 200°C FS and -200 to 1000°C at 1000°C
K: -200 to 200°C at 200°C FS and -200 to 1200°C at 1200°C |
|
Voltage
measurement range |
10, 20, 50, 100, 200, 500 mV, 1, 2,
5, 10, 20, and 50 V FS |
|
Range accuracy |
Temperature: Within 0.5% FS (Within
1% FS at 0°C or lower)
Voltage: Within 0.5% FS |
|
Linearity |
Within 0.1% FS |
|
Allowable input
voltage |
50 V (AC or DC peak values) |
|
CMV |
300V (AC or DC peak values) |
|
Frequency response |
DC to 40 kHz (+0.5, -3 dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -18
dB/OCT) 1, 30, 500, 5 kHz, and OFF (+0.5, -3 dB) |
|
Reference junction
compensation |
Internal/external switchable |
|
A/D converter |
14 bits, 100 kHz max. |
|
Weight |
200 g |
Event Amp (AP11-105)
|
Number of channels |
8 channels/unit |
|
Input signal |
Sets voltage or contact for each
channel
Voltage input: Input voltage range 0 to +24 V
|
Detection level |
H level...2.5 V or higher |
| |
L level...0.5 V or lower |
|
Contact input:
|
Close...250 or lower |
| |
Open...2 k or
higher |
|
|
Response time |
1µs or faster (Input level should be
+5 V or higher.) |
|
Input connector |
Round DIN connector 8P...2 pcs.
Event amp side: XT2B-0800(Conformed to DIN45326) |
|
Weight |
100 g |
2-CH AC Strain Amp (AP11-104)
|
Number of channels |
2 channels/unit, isolated balanced input, NDIS
connector |
|
Applicable gauge
resistance |
120 to
1k |
|
Bridge power
supply |
Sine wave 2 Vrms, 5 kHz (Bridge
power supply is supplied separately.) |
|
Gauge factor |
1.9 to 2.2 |
|
Range of balance |
Resistance: 2% (10000 µ )or
lower Capacitance: 2000 pF or lower |
|
Balance method |
Resistance: Auto-balance
Capacitance: Auto-balance (500 pF or lower, eliminating
constantly)
Balance time: Within 1s at 1 ch Within 0.5%Fs Remained voltage
accuracy |
|
Maximum
sensitivity |
Over full-scale at 500 µ (at
2 V or higher bridge voltage) |
|
Measurement range |
1, 2, 5, 10, 20 kµ FS |
|
Internal
calibrator and accuracy |
0.5 k, 1 k, 2 k, 3 k, 5 kµ Accuracy:
Within 0.5% FS |
|
Linearity |
0.2% FS |
|
CMV |
300 VAC |
|
Frequency response |
DC to 2 kHz (+0.5, -3dB) |
|
Low-pass filter |
Butterworth type (attenuation
factor: -12 dB/OCT)
10, 30 , 100, 300 Hz, and OFF (+0.5, -3 dB) |
|
A/D converter |
16 bits, 100 kHz max. |
|
Temperature
stability |
Zero point: Within 0.05% FS/°C
Gain (range): Within 0.05% FS/°C |
|
Weight |
285 g |
2-CH DC Strain Amp (AP11-110)
|
Number of channels |
2 channels/unit, isolated balanced
input, NDIS connector |
|
Input coupling |
DC |
|
Input impedance |
10M +10M or
higher |
|
Bridge power
supply |
2 V and 5 V |
|
Applicable gauge
resistance |
120 to 2 k
(at BV = 2V),350 to 2k
(at BV=5 V) |
|
Gauge factor |
2 |
|
Range of balance |
2% (15000 µ or
lower |
|
Balance method |
Resistance: Auto-balance
Balance time: 0.5 sec max. at 1ch
Remained voltage accuracy: Within 0.3% FS |
|
Measurement range |
2, 5, 10, 20, 50 kµ FS
(at BV = 2 V)
0.8, 2, 4, 8, 20 kµ FS
(BV = 5 V) |
|
Voltage
measurement range |
2, 5, 10, 20, 50 mV FS |
|
Range accuracy |
Within 0.3% FS |
|
Linearity |
0.1% FS |
|
Allowable input
voltage |
8 V (DC or AC peak values) |
|
Frequency response |
DC to 50 kHz (+0.5, -3dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -12
dB/OCT)
10, 30, 300 Hz, 1 kHz, and OFF (+0.5, -3 dB) |
|
Temperature
stability |
Zero point: Within 0.1% FS/°C
Gain (range): Within 0.01% FS/°C |
|
CMV |
300 VAC |
|
A/D converter |
16 bits, 100 kHz max |
|
Weight |
240 g |
2-CH Vibration/RMS Amp (AP11-109)
|
Input |
2 channels/unit, isolated unbalanced
input, isolated BNC connector |
|
Input coupling |
AC and DC |
|
Input impedance |
10 M or
higher |
|
Power supply for
sensor |
2 mA, 18 V or higher |
|
Measurement range |
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,
100, 200, 500 V |
|
Range accuracy |
DC amp range: Within 0.3% FS (Within
0.8% FS at 500 V)
RMS amp range: Within 2% FS (at DC and 40 Hz to 20 kHz) |
|
Linearity |
Within 0.1% FS |
|
Crest factor |
2.8 max. (RMS amp) |
|
CMV |
Unit only: 42 V (DC or AC peak
values)
When isolated BNC cable (optional) is used: 300 V AC |
|
Frequency response |
DC coupling: DC to 50 kHz (+1, -3
dB)
AC coupling: 1 to 50 kHz (+1, -3 dB) |
|
Low-pass filter |
Butterworth type (attenuation
factor: -24 dB/OCT)
30, 100, 300 Hz, 1 kHz, and OFF |
|
High-pass filter |
Butterworth type (attenuation
factor: -24 dB/OCT)
10, 30, 100 Hz, and OFF |
|
A/D converter |
16 bits, 100 kHz max. |
|
Temperature
stability |
Zero point: Within 0.02% FS/°C
Gain (range): Within 0.01% FS/°C |
|
Weight |
270 g |
2-CH FFT Amp (AP11-102)
|
Input |
2 channels/unit, isolated unbalanced
input, isolated BNC connector |
|
Input coupling |
AC and DC (only AC coupling when
amp-embedded charge accelerometer is used) |
|
Input impedance |
1 M or
higher |
|
Power supply for
sensor |
2 mA, 18 V or higher |
|
Measurement range |
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,
100, 200, 500 V |
|
Range accuracy |
Within 0.3% FS (Within 0.8% FS at
500 V) |
|
Linearity |
Within 0.1% FS |
|
Maximum input
voltage |
500 V (DC or AC peak values) (30 V
at AC coupling of 0.1 to 5 V) |
|
CMV |
Unit only: 42 V (DC or AC peak
values)
When isolated BNC cable (optional) is used: 300 V AC |
|
Frequency response |
DC coupling: DC to 50 kHz (+0.5, -3
dB)
AC coupling: 0.3 to 50 kHz (+0.5, -3 dB) |
|
Low-pass filter |
Bessel type (attenuation factor: -24
dB/OCT)
30, 300 Hz, 3 kHz, and OFF (+0.5, -3 dB) |
|
Anti-aliasing
filter |
20, 40, 80, 200, 400, 800 Hz, 2, 4,
8, 20, and 40 kHz
Drop characteristic: -72 dB typ. at 1.5 x fc |
|
Offset accuracy |
Within 0.3% FS (at 25 °C) |
|
A/D converter |
16 bits, 100 kHz max. |
|
Temperature
stability |
Zero point: Within 0.02% FS/°C
Gain (range): Within 0.01% FS/°C |
|
Weight |
240 g |
F/V Converter (AP11-108)
|
Input |
1 channel/unit, isolated unbalanced
input, isolated BNC connector |
|
Input coupling |
AC and DC |
|
Input impedance |
100 M
or higher |
|
Input frequency
range |
1 Hz to 10 kHz (Pulse width: 20 µs
or longer) |
|
Measurement range |
0.1, 0.2, 0.5, 1, 2, 5, 10 kHz FS |
|
Range accuracy |
Within 0.5% FS (Within 0.8% FS at
500 V) |
|
Linearity |
Within 0.3% FS |
|
Trigger level |
Selectable from 0 V and 2.5 V |
|
Allowable input
voltage |
100 V (AC or DC peak values) |
|
CMV |
Unit only: 42 V (AC or DC peak
values)
Between isolated BNC and safety terminal (optional): 300 VAC
|
|
Response time |
Approx. 20 ms (at the range of 10
kHz) |
|
A/D converter |
16 bits, 100 kHz max. |
|
Temperature
stability |
Zero point: Within 0.03% FS/°C
Gain (range): Within 0.02% FS/°C |
|
Weight |
240 g |
Charge Converter (AP11-901)
|
Gain |
1.0 mV/pC ±5% |
|
Max. input charge |
5000 pC |
|
Frequency range |
Approx. 1.6 Hz to 50 kHz |
|
Max. output
voltage |
5 Vp-p or lower |
|
Drive voltage |
12 to 25 VDC |
|
Drive current |
0.5 to 5.0 mA |
|
Rated noise |
20 µVrms or lower |
|
Phase |
180°(Inverted input is output.) |
|
Operating
temperature |
-20 to 80°C |
|
Connector |
Input: Miniature connector (10-32)
Output: Male BNC terminal |
|
Weight |
20 g |
Charge Converter (AP11-902)
|
Gain |
1.0 mV/pC ±5% |
|
Max. input charge |
5000 pC |
|
Frequency range |
Approx. 1.6 Hz to 50 kHz |
|
Max. output
voltage |
5 Vp-p or lower |
|
Drive voltage |
12 to 25 VDC |
|
Drive current |
0.5 to 5.0 mA |
|
Rated noise |
20 µVrms or lower |
|
Phase |
180°(Inverted input is output.) |
|
Operating
temperature |
-20 to 80°C |
|
Connector |
Input: Miniature connector (10-32)
Output: Male BNC terminal |
|
Weight |
65 g |
Charge Converter (AP11-903)
|
Gain |
1.0 mV/pC ±5% |
|
Max. input charge |
50000 pC |
|
Frequency range |
Approx. 1.6 Hz to 50 kHz |
|
Max. output
voltage |
5 Vp-p or lower |
|
Drive voltage |
12 to 25 VDC |
|
Drive current |
0.5 to 5.0 mA |
|
Rated noise |
20 µVrms or lower |
|
Phase |
180°(Inverted input is output.) |
|
Operating
temperature |
-20 to 80°C |
|
Connector |
Input: Miniature connector (10-32)
Output: Male BNC terminal |
|
Weight |
65 g |

| |
Name |
Model |
Rating |
|
Main unit |
Omniace II |
RA1100 |
Data acquisition system without printer
|
| RA1200 |
Data acquisition system with printer, paper
speed: 25 mm/s max. |
| RA1300 |
Data acquisition system with printer, paper
speed: 100 mm/s max. |
|
Standard accessories |
AC power cable x 1, Recording paper x 1
(RA1200/1300), Paper holder x 1 (RA1200/1300), Slot cover plate
for input and interface units x 1 set, and user's manual x 1 set |
|
OMNIACERA2300

Simple“Pen-recorder”Mode
Visualized amp.setting-display & touch panel provide simple
operation like Pen-Recorder. |
Dynamic wave form in Wide Display
12.1 inches TFT color display improves visibility of measuring
data |
Direct input
from Sensors
11 types of input amp. are available for Voltage, Strain, Temp,
Vibration,Pressure, F/V and other signal input. |
Large size HDD
like 120days-recording
Built-in large capacity 40GB HDD enables 120 days-recording
(based on sampling speed 10ms, 16ch) |
High-speed
thermal printing
Chart speed 100mm/s (recording density : 20 dots/mm) to 1mm/min
(recording density 80 dots/mm) is selectable. |
16 ch Analog +
16 ch Digital
Max. 16ch signal inputs (8 slots) are prepared or various
sensors. And 16ch Digital are available separately from these
input slots at option, for measurement of diverse signal-timing
and contact-condition. |
Standard LAN and
USB I/F
LAN (100BASE-T) for data communication, and USB (1.1) for
connecting to external storage device are equipped as standard
I/F. |
|
Select most suitable measurement mode for your application
5 measurement modes are available, such as Pen-Recorder like
pen-writing recording, HD-Recorder for long term recording.Each
measurement mode is dedicated for each application. OMNIACE III
RA2300 offers simple and certain operation for exact data
acquisition.
 |
 |
|
|
TC/DC
amp (AP11-107)
This amp can be used as a thermocouple amp or DC amp. When
temperature measurement using a thermocouple amp and voltage
measurement using a DC amp are necessary, this amp is useful
because no amp replacement is required.Moreover, four types of
thermocouples, R, T, J, and K, can be connected and high-speed
temperature measurement as fast as 10 µs is possible. This amp
contains one channel per unit.
Event amp (AP11-105)
This amp is a logic amplifier that can input eight channels of
voltage or contact signals. It is useful for measuring timing or
contact signals.
Voltage input: Input voltage 0 to 24 V
H level: 2.5 V or higher
L level: 0.5 V or lower
Contact input:
Open: 2 k
or higher
Short: 250
or lower
2-ch AC strain amp (AP11-104)
This amp is a 2-channel amplifier that inputs signals from
strain gauges or straingauge based sensors. Due to the AC bridge
type, it features high noise resistance(against the noise caused
by the line voltage), high stability, and a high resolution of
16 bits. Low-noise measurement is possible for measurement using
strain gauges, which do not have any shielded wires.
Auto-balance
Bridge voltage: 5 kHz, 2 V rms
Response frequency: 0 to 2 kHz
Gauge factor adjustment function: 1.9 to 2.2
2-ch DC strain amp (AP11-110)
This amp is a 2-channel amplifier that inputs signals from
strain gauges or strain gauge based sensors. Due to the DC
bridge type, it features measurement of high-speed strain
signals at a high resolution of 16 bits.
Auto-balance
Bridge voltage: 2 VDC, 5 VDC
Response frequency: 0 to 50 kHz
2-ch vibration/RMS amp (AP11-109)
This is a 2-channel amplifier that features a constant-current
power supply for amp-embedded sensors, an RMS conversion
function, and a band-pass filter.The RMS function in this amp is
effective for monitoring the basic frequency level fluctuation
in rotation equipment and the fluctuation of the AC power
level.For vibration measurement, an amp-embedded charge sensor
can be directly connected to this amp unit. For voltage
measurement, 500 V (DC or AC peak values)can be directly input.
In addition, the band-pass filter offers filtering of necessary
frequencies to perform the RMS conversion so that clear
waveforms for judgment are available.
Power for sensor: 2 mA/18 V or higher
Voltage measurement range: 100 mV to 500 V
Band-pass filter: -24 dB/oct
A/D: 16 bits
2-ch FFT amp (AP11-102)
This is a 2-channel amplifier that incorporates an anti-aliasing
filter required for the FFT processing and the power for an
amp-embedded sensor (constant current).The anti-aliasing filter
has a dropout feature at -72 dB/1.5 fc so that the aliasing
generated in FFT can be eliminated.
Power for sensor: 2 mA/18 V or higher
Voltage measurement range: 100 mV to 500 V
A/D: 16 bits
Fmax: 40 kHz
F/V converter (AP11-108)
This unit offers frequency-to-voltage conversion, which is
useful in recording rotation changes and revolutions in rotation
equipment.
Input frequency: 10 kHz max.
Response time: 20 ms (Range of 10 kHz) |
2-ch
high-resolution amp (AP11-101)
This is a DC amp that can record at a high resolution of 16
bits. Channels are isolated from each other and have a withstand
voltage for input of 500 V max. (DC or AC peak values). This amp
contains two channels per unit.
|
2-ch high-speed DC
amp (AP11-103)
This is a DC amp that can acquire data at a high-speed sampling
rate of 1 µs.Channels are isolated from each other and have a
withstand voltage for input of 500 V max. (DC or AC peak
values). This amp contains two channels per unit.
|
2-ch TC/DC amp (AP11-106)
This amp can be used as a thermocouple amp or DC amp. When
temperature measurement using a thermocouple amp and voltage
measurement using a DC amp are necessary, this amp is useful
because no amp replacement is required.Moreover, five types of
thermocouples, R, T, J, K, and W can be connected and high-speed
temperature measurement as fast as 10 µs is possible. Channels
are isolated from each other and have an input voltage of 50 V
max. (DC or AC peak values). This amp contains two channels per
unit.
|
|
 |
|
Flaw Detection in
Conveyor Line in Steel Mill |
Operation Test of
Power Switchboards and Production Facilities |
By using the filing function, which operates
synchronously with external signals, data can be acquired for a
long term. Since the RA1000 Series is small and compact, it does
not occupy a large measurement space. |
The RA1000 Series is used to measure the
operation timing of relays and voltage and current changes in
the main power for operation. |
|
Torque Measurement of Engines
or Motors |
Wide Variety of Tests in
Railroad Car |
In automobile testing, measurement of
revolution, strain, pressure, vibration, torque, temperature,
etc is executed. The RA1000 Series incorporates a variety of
amplifiers, which can eliminate the need to install separate
amplifiers and thus enable operators to perform the tests in a
small space.
By using the filing function of the RA1000 Series, long-term
data storage in external media at a high sampling speed (200 µs
max.) is possible. |
The RA1000 Series can accept oil pressure, air
pressure, and acceleration in its AC strain amp, and save the
data in the MO directly. Since the RA1000 Series incorporates
strain amplifiers, external amplifiers are not necessary, which
makes the measurement system compact.
Also, a long-term multi-channel measurement is possible by using
the filing function.
|
|
Governor Test |
Drop Test for Package |
The RA1000 Series is used as a recorder for the
governor test, in which changing signals such as water pressure,
revolution, generator voltage, current, governor, stroke, and
pressure regulation stroke are recorded.
Since the data saved in the memory can be automatically stored
in a PC card, database creation is possible without expending
excessive time and effort. After storing the data, you can
confirm the data on the monitor, which eliminates errors caused
by variations in reading by operators. |
A drop test for packages is performed in order
to avoid the damage that may occur during delivery of packaged
electronic equipment.
Generally, packages contain materials for packing to absorb
shock. This test measures the impact acceleration to obtain the
data that is used for the improvement of materials for packing.
There are two types of accelerometers that can be used in the
drop test: a piezo-electric type or a strain gauge type. The
RA1000 Series supports both types. |
|
Monitoring of Vibration Level
of Rotating Machine |
Magnus Test |
In order to maintain rotating
equipment such as motors, gears, bearings, piezo-electric
accelerometers are used to check their vibration level.In this
case, a band-pass filter that picks up the signals in the
vicinity of the frequency of the rotation is used and RMS
conversion is performed based on the signals so that the level
changes become more visible.Waveform judgment setting is also
possible. |
A Magnus test is a test to examine
the response of a sample of tissue that expands or contracts,
such as the blood vessel, enteron, or trachea of a mouse or rat,
to the medicine applied to the sample.
This test requires a lot of repetition work. The filing and
trigger functions of the RA1000 Series provide automatic
processing of this work.Moreover, since strain amps can be
incorporated, a tension sensor can be directly connected to the
main unit,thus eliminating troublesome cabling.In addition, an
operator can examine waveforms on the monitor during data
acquisition and readout current values calculated by physical
conversion. |
|